Chapter 1 Performance & Specifications RIGOL
Service Guide for DM3058
1-3
Electric Technical Specifications
DC Characteristics
Accuracy Specifications (% of reading + % of range)
[1]
Function
Range
[2]
Test Current
or Burden
Voltage
1 Year
23 ±5
Temperature
Coefficient
0 to 18
28 to 55
DC Voltage
200.000 mV
0.015 + 0.004
0.0015 + 0.0005
2.00000 V
0.015 + 0.003
0.0010 + 0.0005
20.0000 V
0.015 + 0.004
0.0020 + 0.0005
200.000 V
0.015 + 0.003
0.0015 + 0.0005
1000.00 V
[4]
0.015 + 0.003
0.0015 + 0.0005
DC Current
[8]
200.000 μA
<8 mV
0.055 + 0.005
0.003 + 0.001
2.00000 mA
<80 mV
0.055 + 0.005
0.002 + 0.001
20.0000 mA
<0.05 V
0.095 + 0.020
0.008 + 0.001
200.000 mA
<0.5 V
0.070 + 0.008
0.005 + 0.001
2.00000 A
<0.1 V
0.170 + 0.020
0.013 + 0.001
10.0000 A
[5]
<0.3 V
0.250 + 0.010
0.008 + 0.001
Resistance
[3]
200.000 Ω
1 mA
0.030 + 0.005
0.0030 + 0.0006
2.00000
1 mA
0.020 + 0.003
0.0030 + 0.0005
20.0000
100 μA
0.020 + 0.003
0.0030 + 0.0005
200.000
10 μA
0.020 + 0.003
0.0030 + 0.0005
2.00000
1 μA
0.040 + 0.004
0.0040 + 0.0005
10.0000
200nA
0.250 + 0.003
0.0100 + 0.0005
100.000
200 nA || 10
1.75 + 0.004
0.2000 + 0.0005
Diode Test
2.0000 V
[6]
1 mA
0.05 + 0.01
0.0050 + 0.0005
Continuity
2000 Ω
1 mA
0.05 + 0.01
0.0050 + 0.0005
Remarks:
[1] Specifications are for 0.5 hour warm-up, Slow measure and calibration temperature 18 ~
28 .
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specifications are for 4-wire measure or 2-wire measure under REF operation. ± 0.2 Ω of
extra errors will be generated when perform 2-wire measure without REF operation.
[4] Plus 0.02 mV of error per 1V after the first ± 500 VDC.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current higher than DC
7 A or AC RMS 7 A.
[6] Accuracy specifications are only for Input terminal measuring. The typical value of current
under measure is 1 mA. Voltage drop at the diode junction may vary with changes of current
supply.
DC Voltage
Input Resistance: 200 mV and 2 V range 10 or >10 selectable
(Input signals which exceed ± 2.5 V in these ranges will pass the 100
(typical) clamp resistance).
20 V, 200 V and 1000 V range 10 MΩ±2%
Input offset current: <90 pA, 25