Chapter 1 Performance & Specifications RIGOL
Service Guide for DM3058
Electric Technical Specifications
DC Characteristics
Accuracy Specifications (% of reading + % of range)
[1]
Test Current
or Burden
Voltage
Temperature
Coefficient
0 ℃ to 18 ℃
28 ℃ to 55 ℃
Remarks:
[1] Specifications are for 0.5 hour warm-up, “Slow” measure and calibration temperature 18 ℃ ~
28 ℃.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specifications are for 4-wire measure or 2-wire measure under “REF” operation. ± 0.2 Ω of
extra errors will be generated when perform 2-wire measure without “REF” operation.
[4] Plus 0.02 mV of error per 1V after the first ± 500 VDC.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current higher than DC
7 A or AC RMS 7 A.
[6] Accuracy specifications are only for Input terminal measuring. The typical value of current
under measure is 1 mA. Voltage drop at the diode junction may vary with changes of current
supply.
DC Voltage
Input Resistance: 200 mV and 2 V range 10 MΩ or >10 GΩ selectable
(Input signals which exceed ± 2.5 V in these ranges will pass the 100 kΩ
(typical) clamp resistance).
20 V, 200 V and 1000 V range 10 MΩ±2%
Input offset current: <90 pA, 25 ℃