RIGOL Performance Verification Guide DG1000Z Series Function/Arbitrary Waveform Generator Mar. 2014 RIGOL Technologies, In
RIGOL DG1000Z Performance Verification Guide VIII Content Conventions in this Manual: DG1000Z series function/arbitrary waveform generator includes
Contents RIGOL DG1000Z Performance Verification Guide IX Contents Guaranty and Declaration ...
Chapter 1 Test Overview RIGOL DG1000Z Performance Verification Guide 1-1 Chapter 1 Test Overview Test Preparations Before performing the test, mak
RIGOL Chapter 1 Test Overview DG1000Z Performance Verification Guide 1-2 50Ω Load 50Ω/1W -- Power Sensor -30dBm to +20dBm Agilent N8482A Power Se
Chapter 1 Test Overview RIGOL DG1000Z Performance Verification Guide 1-3 Technical Parameters Chapter 2 of this manual provides the corresponding sp
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-1 Chapter 2 Performance Verification Test This chapter int
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-2 Frequency Accuracy Test Specification: Frequency characteri
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-3 5. Record the reading of the frequency counter and judge wh
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-4 AC Amplitude Accuracy Test Specification: Output Characteri
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-5 b) Set the output waveform of CH1 to a sine waveform with 1
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-6 Test Record Form: Amplitude Setting Value Setting Measureme
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-7 DC Offset Accuracy Test Specification: Output Characteristi
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-8 respectively and judge whether the readings are within the s
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-9 AC Flatness Test Specification: Output Characteristic Flatn
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-10 the multimeter. Wherein, readingV is the measurement value
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-11 amplitude. Set the frequency factor of the power meter to 1
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-12 Harmonic Distortion Test Specification: Sine Wave Spectrum
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-13 4. Turn on and set the spectrum analyzer: a) Set the refe
RIGOL DG1000Z Performance Verification Guide I Guaranty and Declaration Copyright © 2014 RIGOL Technologies, Inc. All Rights Reserved. Trademark I
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-14 measurement value For example, when the output waveform fre
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-15 Spurious Signal Test Specification: Sine Wave Spectrum Pur
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-16 5. After the spectrum analyzer finishes a sweep, use Peak
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-17 14. Keep the reference level, input attenuation, resolutio
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-18 Rise/Fall Time Test Specification: Signal Characteristic S
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-19 4. Turn on and set the oscilloscope: a) Set the vertical
RIGOL Chapter 2 Performance Verification Test DG1000Z Performance Verification Guide 2-20 Overshoot Test Specification: Signal Characteristic Square
Chapter 2 Performance Verification Test RIGOL DG1000Z Performance Verification Guide 2-21 6. Repeat steps 1 to 5 to test the overshoot of CH2 and re
Appendix RIGOL DG1000Z Performance Verification Guide 1 Appendix Appendix A: Test Result Record Form RIGOL DG1000Z Series Function/Arbitrary Wa
RIGOL DG1000Z Performance Verification Guide II General Safety Summary Please review the following safety precautions carefully before putting the
RIGOL Appendix DG1000Z Performance Verification Guide 2 DC Offset Accuracy Test Offset Setting Value Setting Measurement Value Specification Pass/Fa
Appendix RIGOL DG1000Z Performance Verification Guide 3 Harmonic Distortion Test Frequency Setting Value Setting Measurement Value Calculation Resu
RIGOL Appendix DG1000Z Performance Verification Guide 4 Rise/Fall Time Test Waveform Setting Measurement Value Specification Pass/Fail Square Fre
Appendix RIGOL DG1000Z Performance Verification Guide 5 Channel: CH2 Frequency Accuracy Test Waveform Setting Value Measurement Value Specificatio
RIGOL Appendix DG1000Z Performance Verification Guide 6 DC Offset Accuracy Test Offset Setting Value Setting Measurement Value Specification Pass/Fa
Appendix RIGOL DG1000Z Performance Verification Guide 7 Harmonic Distortion Test Frequency Setting Value Setting Measurement Value Calculation Resu
RIGOL Appendix DG1000Z Performance Verification Guide 8 Rise/Fall Time Test Waveform Setting Measurement Value Specification Pass/Fail Square Fre
Appendix RIGOL DG1000Z Performance Verification Guide 9 Appendix B: Performance Specifications Unless otherwise specified, all the specifications
RIGOL Appendix DG1000Z Performance Verification Guide 10 Sine Wave Spectrum Purity Harmonic Distortion Typical (0dBm) DC-10MHz (included): <-65dBc
Appendix RIGOL DG1000Z Performance Verification Guide 11 Overshoot Typical (1Vpp) ≤5% Jitter (rms) Typical (1Vpp) ≤5MHz: 2ppm+200ps >5MHz: 200ps
RIGOL DG1000Z Performance Verification Guide III Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powere
RIGOL Appendix DG1000Z Performance Verification Guide 12 Accuracy ±(1% of setting+5mV+0.5% of amplitude) Waveform Output Impedance 50Ω (typical) Pr
Appendix RIGOL DG1000Z Performance Verification Guide 13 PSK Carrier Waveform Sine, Square, Ramp, Arb (except DC) Source Internal/External Modula
RIGOL Appendix DG1000Z Performance Verification Guide 14 Hold/Return Time 0ms to 500s Trigger Source Internal, External, Manual Mark Falling edge
Appendix RIGOL DG1000Z Performance Verification Guide 15 GateTime4 1.342s GateTime5 10.73s GateTime6 >10s Trigger Characteristics Trigger
RIGOL Appendix DG1000Z Performance Verification Guide 16 Sync Output Level TTL-compatible Impedance 50Ω, nominal value Overvoltage Protection Over
Appendix RIGOL DG1000Z Performance Verification Guide 17 Weight without package: 3.2kg with package: 4.5kg Interfaces USB Host, USB Device, LAN IP
RIGOL DG1000Z Performance Verification Guide IV Handling Safety. Please handle with care during transportation to avoid damages to buttons, knob int
RIGOL DG1000Z Performance Verification Guide V Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPerso
RIGOL DG1000Z Performance Verification Guide VI Vermeiden Sie ungeschützte Verbindungen. Berühren Sie keine unisolierten Verbindungen oder Baugruppe
RIGOL DG1000Z Performance Verification Guide VII Document Overview This manual is used to guide users to correctly test the performance specification