Chapter 2 Front Panel Operation RIGOL
User’s Guide for DSA800 Series
2-59
Chan Pwr
Measurement Interface:
Figure 2-16 Channel Power Measurement Interface
Measurement Results: channel power and power spectral density.
Channel Power: power within the integration bandwidth.
Power Spectral Density: power (in dBm/Hz) normalized to 1 Hz within the
integration bandwidth.
Measurement Parameters: average number, average mode, integration
bandwidth and channel power span.
1. Avg Num
Specify the number of averages used when calculating the measurement
result and the default is Off. You can use the numeric keys, knob or direction
keys to modify this parameter.