Trigger Modes
The oscilloscope provides seven trigger modes: Edge, Pulse Width, Slope, Video,
Alternate, Pattern (only for DS1000D series) and Duration trigger (only for DS1000D
series).
Edge: An edge trigger occurs when the trigger input passes through a specified
voltage level at the specified slope direction.
Pulse Width: Use this trigger type to catch pulses with certain pulse width.
Slope: The oscilloscope begins to trigger according to the signal rising or falling
speed.
Video: Use video trigger on fields or lines for standard video signals.
Alternate: Trigger on non-synchronized signals of the dual channels.
Pattern: To Trigger through detecting a specified code.
Duration: To trigger within a specified time on the conditions of a specified code.