Pass/Fail connection
The Pass/Fail connection has a optically isolated output. Users need to connect to
another circuit to fulfill the function.
Before connecting external circuit, make sure the maximum voltage/current do not
exceed the internal photoMOS relay’s rating, 400V/100mA. DS1000 series adopts the
technique of Optical Isolation, output device has no polarity and can be used in any
circuit within the ratings.
Figure 2-99 Schematic Diagram of Pass/Fail output