AC Characteristics
Accuracy Specifications (% of reading + % of range)
[1]
Function Range
[2]
Frequency Range
1 Year
23±5
Temperature
Coefficient
0 to 18
28 to 55
10Hz-45Hz 1.0 + 0.1 0.02+0.02
45Hz-20kHz 0.2 + 0.1 0.02+0.02
20kHz-50kHz 2.0 + 0.2 0.02+0.02
200.000mV
50kHz-100kHz 4.0 + 0.2 0.02+0.02
10Hz-45Hz 1.0 + 0.1 0.02+0.02
45Hz-20kHz 0.2 + 0.1 0.02+0.02
20kHz-50kHz 1.0 + 0.1 0.02+0.02
True RMS AC
Voltage
[3]
2V to
750.00V
50kHz-100kHz 2.0 + 0.2 0.02+0.02
10Hz-45Hz 1.5+0.1 0.02+0.02
45Hz-2kHz 0.5+0.1 0.02+0.02
20.0000mA
2kHz-10kHz 2.0+0.2 0.02+0.02
10Hz-45Hz 1.5+0.1 0.02+0.02
45Hz-2kHz 0.5+0.1 0.02+0.02
200.000mA
2kHz-10kHz 2.0+0.2 0.02+0.02
10Hz-45Hz 1.5+0.5 0.02+0.05
45Hz-2kHz 0.5+0.5 0.02+0.05
1.00000A
2kHz-10kHz 2.0+0.5 0.02+0.05
10Hz-45Hz 1.5+0.1 0.02+0.02
45Hz-2kHz 0.5+0.1 0.02+0.02
True RMS AC
Current
[4, 6]
10.0000A[6]
2kHz-5kHz 2.0+0.2 0.02+0.02