AC Characteristics
Accuracy Specifications (% of reading + % of range)
[1]
Function
Range
[2]
Frequency Range
1 Year
23±5
Temperature
Coefficient
0 to 18
28 to 55
True RMS AC
Voltage
[3]
200.000mV
10Hz-45Hz
1.0 + 0.1
0.02+0.02
45Hz-20kHz
0.2 + 0.1
0.02+0.02
20kHz-50kHz
2.0 + 0.2
0.02+0.02
50kHz-100kHz
4.0 + 0.2
0.02+0.02
2V to
750.00V
10Hz-45Hz
1.0 + 0.1
0.02+0.02
45Hz-20kHz
0.2 + 0.1
0.02+0.02
20kHz-50kHz
1.0 + 0.1
0.02+0.02
50kHz-100kHz
2.0 + 0.2
0.02+0.02
True RMS AC
Current
[4, 6]
20.0000mA
10Hz-45Hz
1.5+0.1
0.02+0.02
45Hz-2kHz
0.5+0.1
0.02+0.02
2kHz-10kHz
2.0+0.2
0.02+0.02
200.000mA
10Hz-45Hz
1.5+0.1
0.02+0.02
45Hz-2kHz
0.5+0.1
0.02+0.02
2kHz-10kHz
2.0+0.2
0.02+0.02
1.00000A
10Hz-45Hz
1.5+0.5
0.02+0.05
45Hz-2kHz
0.5+0.5
0.02+0.05
2kHz-10kHz
2.0+0.5
0.02+0.05
10.0000A[6]
10Hz-45Hz
1.5+0.1
0.02+0.02
45Hz-2kHz
0.5+0.1
0.02+0.02
2kHz-5kHz
2.0+0.2
0.02+0.02