Command Systems RIGOL
Programming Guide for DS1000E, DS1000D Series
2-17
:TIM:OFFS? The query returns 1.000e+00.
3. :TIMebase[:DELayed]:SCALe
Command Format:
:TIMebase[:DELayed]:SCALe <scale_val>
:TIMebase[:DELayed]:SCALe?
Function Explanation:
The commands set and query the horizontal scale for MAIN or DELayed
timebase, the unit is s/div (seconds/grid), thereinto:
In YT mode, the range of <scale_val> is 2ns - 50s;
In ROLL mode, the range of <scale_val> is 500ms - 50s;
In MAIN state, the item [:DELayed] should be omitted.
Returned Format:
The query returns the setting value of <scale_val> in s.
Example:
:TIM:MODE MAIN Setup the timebase as MAIN.
:TIM:SCAL 2 Setup its scale as 2s.
:TIM:SCAL? The query returns 2.000e+00.
4. :TIMebase:FORMat
Command Format:
:TIMebase:FORMat <value>
:TIMebase:FORMat?
Function Explanation:
The commands set and query the horizontal timebase. <value> could be XY, YT
or SCANning.
Returned Format:
The query returns X-Y, Y-T or SCANNING.
Example: