Command Systems RIGOL
Programming Guide for DS1000B Series
2-63
16. :MEASure:PWIDth?
Command Format:
:MEASure:PWIDth? [<source>]
Function:
The command measures the positive pulse width of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 5.000e-004, and the unit is s.
17. :MEASure:NWIDth?
Command Format:
:MEASure:NWIDth? [<source>]
Function:
The command measures the negative pulse width of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 5.000e-004, and the unit is s.
18. :MEASure:PDUTycycle?
Command Format:
:MEASure:PDUTycycle? [<source>]
Function:
The command measures the positive duty cycle of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 5.000e001, and the unit is %.