Rigol DM3058 Specifications Page 186

  • Download
  • Add to my manuals
  • Print
  • Page
    / 264
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 185
RIGOL
Programming Guide for DM3058
The command sets or queries the enable state of 2WR NULL measuring function.
Explanations:
[SENSe:]RESistance:NULL[:STATe]{ON|OFF}: Enable or disable 2WR NULL
measuring function.
[SENSe:]RESistance:NULL[:STATe]?: Query the enable state of 2WR NULL
measuring function.
The command [SENSe:]RESistance:NULL:VALue can be used to set NULL value.
The commands CONFigure:RESistance and MEASure:RESistance? can be
used to disable NULL function (OFF).
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The command returns the enable state of 2WR NULL measuring function.
37. [SENSe:]RESistance:NULL:VALue
Command Format:
[SENSe:]RESistance:NULL:VALue{<value>|MIN|MAX}
[SENSe:]RESistance:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of 2WR measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]RESistance:NULL:VALue{<value>|MIN|MAX}: Set the NULL value of
2WR measurement. MIN is -120% max range of current function, MAX is 120%
max range of current function, DEF = 0.
[SENSe:]RESistance:NULL:VALue?[{MIN|MAX}]: Query the NULL value of 2WR
measurement.
Before sending the command, do run the command [SENSe:]RESistance:NULL
[:STATe]ON to enable NULL function.